Fane: Lunge_30mAs 

Routine:

mA: CAREDose4D

kV: 110

Scan time: 17,39 s

Delay: 3 s

Slice: 5,0 mm

Caudocranial

 

 

 

 

 

 

 

 

 

 

 

Fane: Lunge_30mAs

Scan:

Quality ref. mAs: 30

CARE Dose type:AEC+DOM

kV: 110

Scan time: 17,39 s

Rotation time: 0,6 s

Delay: 3 s

Slice: 5,0 mm

Direction: Caudocranial

Pitch: 1,5

Scan start: Start button

API: None

 

 

 

 

Fane: Lunge_30mAs

Recon:

Kernel: B08s SPECT AC

Window: Lung

Extended FoV: Vinket af

FoV: 700 mm

Series description: AC Lunge 30mAs 5.0 eFoV

Recon job type: Axial

Image order: Craniocaudal

Recon increment: 3,0 mm

 

 

 

 

 

Fane: Lunge_30mAs 

Auto tasking

Recon job: 1

Axial1: SPECT^Lunge_30mAs (Adult),B08s SPECT AC for ECT,2,4

Axial2: SPECT^Lunge_30mAs (Adult),I31s medium smooth +,2,5

Auto recon: Vinket af

Auto postprocessing: None

Requested Procedure: None

Body part examined: Lung